The Informant™ 2 
 
 
 
0019-9211 Rev 9 
 
NOTE:  For dual detectors, the red probe tip will not screw onto a 
probe that has a refrigerant sensor installed. This feature 
safeguards against the possibility of an operator mistakenly thinking 
that a detector with a refrigerant sensor is configured to detect 
combustibles. See the WARNING below. 
 
2.3.  Refrigerant Sensor Installation 
 
Install the refrigerant sensor, filter, and “blue” probe tip onto the end of 
the detector’s flexible probe as shown in Figure 3a. Observe the blue 
color coding on both the sensor and probe tip. 
 
 
WARNING:  Explosion hazard. DO NOT use the Informant 2 with 
its refrigerant sensor installed in an area that is classified by the 
National Electrical Code as being a hazardous location where a 
potential for explosion or fire exist because of flammable gases or 
vapors in the atmosphere. 
 
 
COMBUSTIBLE REFRIGERANT GAS WARNING:  For your 
safety, DO NOT use the Informant 2 with the refrigerant sensor to 
detect refrigerants which are rated as combustible/flammable 
gases  (e.g. ASHRAE - A2 or A3 rated refrigerants). Use with 
combustible/flammable refrigerant gases can cause an explosion 
resulting in death or severe injury. 
 
 
Figure 3a. Sensor Installation (Refrigerant Sensor)