Operation of Specific Models 
SCFF51OSWH MED Series 
These frost-free all-freezers can be built-in or free-standing. They do not require defrosting 
and have a digital thermostat, LED lighting and a lock. At the top of the unit is a cooling 
system that uses a fan to provide cold air. For best operation, do not block the air flow with 
contents.  
FF6L, FF7L, FF28LWH, FF511L, FF521BL, FF590SSHH, FFAR2L, FFAR10, FFAR21L, 
FFAR24L, SCR450L, SCR500BL, SCR600L, SCR610BL PLUS & MED Series;  
SP6DS2D7, SP6DS7, SP6DSSTB7 MED Series only 
These models are all-refrigerators  (no freezer compartment) with automatic defrost. All 
have cold plates on the rear wall, and the coldest spots in the refrigerator will be in the rear 
and toward the bottom. All have interior fans to promote even cooling. In the rear bottom 
above the step, there is a drain line to allow defrost water to exit the refrigerator. Keep this 
drain line clear, or icing inside the refrigerator will occur. FF7, FF511L, FF521BL, and SCR 
series models have lights that are not door-activated, and are controlled by a rocker switch 
on the top right. The light generates a small amount of heat, so minimize the use of the light 
for best operation and energy savings. FF6 models and FFAR10 models all have door-
activated lights. Smaller models and drawer models do not have interior lights.  
NOTE: FF590, models with BI in the model number, and all drawer units are designed for 
built-in operation and need very little clearance on the sides and  above the units. All 
other models
FCL44, FS20L, FS21L, FS22L, FS24L, FS30L, FS60ML, FS407L, FS408BL, VT65ML 
PLUS & MED Series 
  require  at least  1 inch of airflow on the sides and above the units. All 
models operate best when 2  inches  is provided between the rear wall and the back of 
the unit.  
Model FCL44 is a chest freezer; all of the others are upright. These are manual defrost all-
freezers. No internal fans are present, and ice will build up along the cooling coils. 
Periodically defrost the ice (recommended when the ice level reaches a thickness of 1/4 to 
1/2 inch). NEVER USE A METAL INSTRUMENT as you can damage the cooling coils, 
which will void your warranty and often cannot be repaired. To defrost, disconnect from the 
power supply, empty the contents, provide a tray in the bottom to collect water, and use a 
hairdryer or other source of warm air to melt the ice (or leave the door open and the power 
off for 24 to 48 hours). For built-in (BI) models and clearance information, see NOTE 
above.
 
CP35LL PLUS & MED Series; CT66 PLUS Series only 
These models are cycle defrost refrigerator-freezers. The freezers are designed to operate 
at 0°F (-18°C), suitable for medical or laboratory uses. These  models have dual 
evaporators to maintain separate temperatures in the freezer and refrigerator. Defrost the 
freezer carefully, as directed above for manual defrost. The refrigerator section operates in 
the normal range (35 to 45°F, or 2 to 7°C) and defrosts automatically. For built-in (BI) 
models and clearance information, see NOTE above.
 
CM411L, CM421BL, S19LWH PLUS Series 
These models are manual defrost refrigerator-freezers. The freezers in these models are 
designed to operate at 25°F (-4°C) to keep ice frozen, and not for medication or products 
needing colder temperatures. Defrost carefully, following directions above. The refrigerator 
section will operate in the normal range (35 to 45°F, or 2 to 7°C) and is suitable for most 