8.2  AUTO CALIBRATION FOR ALL OTHER MODELS 
 
•  Before entering calibration mode, press [On/Off] to turn the power 
on for 1 minute. 
 
Enter calibration mode 
 
• Press [On/Off] to turn the power off.  
 
• Press [Unit] first and then press [On/Off], while keeping the [Unit] 
key pressed. The display will show the last calibration value used 
(For example, “ L – 1000”), press [Tare] to choose between the two 
calibration weight values. When you have made your selection press 
[Unit] and the display will show the A/D value. 
 
Calibration  
 
•  Wait for the stable indicator to be displayed,  
 
• Press [Unit] -the display will show "00SAVE". After 2 seconds, the 
display will return to A/D value.  
 
•  Place the required weight on the platform. This is the weight that you 
selected as your calibration weight value earlier.  
 
•  After the stable indicator is displayed press [Unit] 
 
•  The display will show "00SAVE". The calibration in complete now.  
 
• Press [On/Off] to turn the power off. Press [On/Off] to turn on the 
power again,  
 
•  Place full weight (eg.2000g for AQT2000) on the platform to verify 
weighing is correct. If not, repeat the calibration procedure
.
 
 
9.0 ERROR 
MESSAGES 
 
EEEEHI Overload 
 
EEEELO 
A/D value too low 
 
 
Low battery 
 
 
Note: 
Calibration may be required before weighing. Read section 8.0 on 
"CALIBRATION" first and if necessary calibrate your balance for accurate 
weighing. 
 