Service Procedures  
Appendix D
AC Amplitude Adjustment (High-Z)
The function generator stores twenty-two calibration constants related to 
HIGH Z output, and sixteen calibration constants related to 50 
Ω output. The 
constants are calculated from the adjustment value entered. The calibration 
constants are stored following completion of SETUP 22 (HIGH Z output) 
and the calibration procedure may be aborted after that point. No calibration 
constants are stored if the procedures are aborted at any other setup. 
1. Use a DMM to measure the function generator ACrms output voltage 
for SETUP 08. These adjustments use a HIGH Z output termination.
2. Send the CAL:VALue <measured Vac> to the function generator.
3. Send CAL? command to the function generator to initiate calibration.
4. Repeat steps 1, 2 and 3 for SETUP 09 through 29.
5. Perform the “AC Amplitude Verification” on page 169.
Table D-13. AC Amplitude (High Z) Adjustment Setups
Nominal Output
SETUP
FREQUENCY
AMPLITUDE
Adjustment for:
08
1 kHz
5.5 V rms
2 dB Output Attenuator
09
1 kHz
4.4 V rms
4 dB Output Attenuator
10
1 kHz
3.5 V rms
6 dB Output Attenuator
11
1 kHz
2.8 V rms
8 dB Output Attenuator
12
1 kHz
2.2 V rms
10 dB Output Attenuator
13
1 kHz
1.7 V rms
12 dB Output Attenuator
14
1 kHz
1.4 V rms
14 dB Output Attenuator
15
1 kHz
1.1 V rms
16 dB Output Attenuator
16
1 kHz
0.88 V rms
18 dB Output Attenuator
17
1 kHz
0.70 V rms
20 dB Output Attenuator
18
1 kHz
0.55 V rms
22 dB Output Attenuator
19
1 kHz
0.44 V rms
24 dB Output Attenuator
20
1 kHz
0.35 V rms
26 dB Output Attenuator
21
1 kHz
0.28 V rms
28 dB Output Attenuator
22
1 kHz
0.22 V rms
30 dB Output Attenuator
23
1 kHz
 5.5 V rms
2 dB Pre-attenuator
24
1 kHz
 4.4 V rms
4 dB Pre-attenuator
25
1 kHz
 3.5 V rms
6 dB Pre-attenuator
26
1 kHz
 2.8 V rms
8 dB Pre-attenuator
27
1 kHz
 2.2V rms
10 dB Pre-attenuator
28
1 kHz
 1.7 V rms
12 dB Pre-attenuator
29
1 kHz
 1.4 Vrms
14 dB Pre-attenuator