Maintenance
Performance Test
6
6-13
3.
 
Verify Accuracy  Configure the 5700A for the output values below and
verify the Spy window measurement is between the minimum and maximum
values. Change the channel 1 range as required (see Step 1).
Volts DC Range
5700A Output
Minimum Reading
Maximum Reading
90 mV
Short Circuit (Zero)
-0.000023V
+0.000023V
90 mV
+90 mV
+0.089965V
+0.090035V
90 mV
-90 mV
-0.090035V
-0.089965V
300 mV
Short Circuit (Zero)
-0.00005V
+0.00005V
300 mV
+300 mV
+0.29991V
+0.30009V
300 mV
-300 mV
-0.30009V
-0.29991V
3V
Short Circuit (Zero)
-0.0004V
+0.0004V
3V
+3V
+2.9992V
+3.0008V
3V
-3V
-3.0008V
-2.9992V
30V
Short Circuit (Zero)
-0.005V
+0.005V
30V
+30V
+29.991V
+30.009V
30V
-30V
-30.009V
-29.991V
50V
Short Circuit (Zero)
-0.04V
+0.04V
50V
+50V
+49.95V
+50.05V
50V
-50V
-50.05V
-49.95V
4.
 
Close Spy Window.
Volts AC Accuracy Test
6-11.
Complete the following procedure to test the accuracy of the volts ac function for
both the 2640A and 2645A. Measurement accuracy applies to all channels, not
just the channel used for the test.
1.
 
Configure Channel 1 for Volts AC  In NetDAQ Logger, configure channel 1
for volts ac, 300 mV range. (See “Configuring Analog Channel Functions" in
Chapter 3 of this manual.)
2.
 
Open Spy Window  Select the Spy command from the Utilities menu. Select
channel 0101 (instrument 01, channel 01) from the Channel list. Click OK to
open the Spy window.