Справочник Пользователя для Fluke 2640A

Скачать
Страница из 353
2640A/2645A NetDAQ
Users Manual
A-4
Table A-2. 2640A/2645A Environmental Specifications
Specification
Characteristic
Warm-up Time
1 hour to rated specifications -or- 15 minutes if relative
humidity (non-condensing) is 50% or less.
Operating Temperature
-20
°
C to 60
°
C (-4
°
F to 140
°
F)
Storage Temperature
-40
°
C to +70
°
C (-40
°
F to +158
°
F)
Relative Humidity
90% maximum for -10
°
C to 28
°
C (14
°
F to 82.4
°
F)
75% maximum for 28
°
C to 35
°
C (82.4
°
F to 95
°
F)
50% maximum for 35
°
C to 60
°
C (95
°
F to 140
°
F)
(3 M
Ω
 range, reduce humidity rating by 25% for 1 hour
warm-up. 3 M
Ω
 range meets full humidity ratings with 2
hour warm-up.)
Altitude
Operating: 2,000m (6,561 ft) maximum
Non-operating: 12,200m (40,000 ft) maximum
Vibration
0.7g at 15 Hz
1.3g at 25 Hz
3g at 55 Hz
Shock
30g half sine per Mil-T-28800E
Bench handling per Mil-T-28800E
2640A/2645A Input/Output Capabilities
A-5.
The following specifications include the input/output functions, including the
Digital I/O, Trigger Out, Trigger In, and Master Alarm output.
Digital I/O
A-6.
Table A-3 provides a summary of the Digital I/O specifications for the 8 Digital
I/O lines (0 to 7). Digital I/O is located on the DIGITAL I/O connector, terminals
0 to 7, and GND.
Table A-3. 2640A/2645A DIGITAL I/O Specification
Specification
Characteristic
Maximum Input Voltage
30V
Minimum Input Voltage
-4V
Isolation
None (dc coupled)
Threshold
1.4V
Hysteresis
500 mV