Specifications
2645A Specifications
A
A-19
2645A AC Voltage Measurement Specifications
A-23.
Tables A-27 to A-29 provide 2645A specifications for the ac voltage measurement
function.
function.
Table A-27. 2645A AC Voltage General Specifications
Specification
Characteristic
Input Impedance
1 M
Ω
in parallel with 100 pF
Maximum Crest Factor
3.0 Maximum
2.0 for rated accuracy
2.0 for rated accuracy
Crest Factor Error
For non-sinusoidal input signals with crest factors
between 2 and 3 and pulse widths
between 2 and 3 and pulse widths
≥
100 µs, add
0.2% to the accuracy specifications.
Common Mode Rejection
80dB minimum at dc, 50 Hz/60 Hz
±
0.1%, 1 k
Ω
imbalance, Slow Rate
Maximum Input Voltage
The lesser voltage of:
30V ac rms from any input terminal to earth.
30V ac rms from any input terminal to earth.
30V ac rms from any terminal input to any other
input terminal.
input terminal.
Maximum Volt-Hertz Product
2x10
6
Volt-Hertz product on any range, normal
mode input.
1x10
6
Volt-Hertz product on any range, common
mode input.
Temperature Coefficient
Linear interpolation between 2 applicable points for
temperatures between 28
temperatures between 28
º
C and 60
º
C, or -10
º
C
and 18
º
C, e.g., if the applicable specification at
28
º
C is 2% and the specification at 60
º
C is 3%,
then the specification at 40
º
C is
(3%-2%)x(40-28)/(60-28)+2%=2.375%.
Accuracy at -20
°
C
Multiply the -10
°
C to + 60
°
C accuracy specification
by 2. After 1 hour warm-up. For accuracy between
-10
-10
°
C and -20
°
C, interpolate linearly.
DC Component Error
The presence of a dc voltage will cause an
indeterminate error in the reading of the ac voltage
on the input.
indeterminate error in the reading of the ac voltage
on the input.