Specifications
2645A Specifications
A
A-19
2645A AC Voltage Measurement Specifications
A-23.
Tables A-27 to A-29 provide 2645A specifications for the ac voltage measurement
function.
Table A-27. 2645A AC Voltage General Specifications
Specification
Characteristic
Input Impedance
1 M
Ω
 in parallel with 100 pF
Maximum Crest Factor
3.0 Maximum
2.0 for rated accuracy
Crest Factor Error
For non-sinusoidal input signals with crest factors
between 2 and 3 and pulse widths 
100 µs, add
0.2% to the accuracy specifications.
Common Mode Rejection
80dB minimum at dc, 50 Hz/60 Hz 
±
0.1%, 1 k
Ω
imbalance, Slow Rate
Maximum Input Voltage
The lesser voltage of:
30V ac rms from any input terminal to earth.
30V ac rms from any terminal input to any other
input terminal.
Maximum Volt-Hertz Product
2x10
6
 Volt-Hertz product on any range, normal
mode input.
1x10
6
 Volt-Hertz product on any range, common
mode input.
Temperature Coefficient
Linear interpolation between 2 applicable points for
temperatures between 28
º
C and 60
º
C, or -10
º
C
and 18
º
C, e.g., if the applicable specification at
28
º
C is 2% and the specification at 60
º
C is 3%,
then the specification at 40
º
C is
(3%-2%)x(40-28)/(60-28)+2%=2.375%.
Accuracy at -20
°
C
Multiply the -10
°
C to + 60
°
C accuracy specification
by 2. After 1 hour warm-up. For accuracy between
-10
°
C and -20
°
C, interpolate linearly.
DC Component Error
The presence of a dc voltage will cause an
indeterminate error in the reading of the ac voltage
on the input.