Instruction Manual
IM-106-4050 Original Issue
March 2006
OxyBalance Oxygen Display 
and Averaging System
Alarm
There are two divisions of the Alarm tab: Active Alarms and Alarm History, 
shown in Figure 3-9 and Figure 3-10.
There are four possible reports that can be displayed.
Average # 1-4 warning
This appears when one of the probes fails or is in calibration. When this Alarm 
appears, the average channel will still display an average O
2
 % for the 
remaining probes in the group.
Average # 1-4 failed
This appears when less than 2 probes in the group are transmitting data for 
the average calculation. No average percentage will be calculated.
Probe # 1-8 calibration in progress
This means that the probe is calibrating and will automatically join its specified 
group when the calibration is complete. Probes in calibration are not factored 
into the channel average.
Probe #1-8 failed
This means the probe has failed and needs to be checked. Refer to Section 6, 
Troubleshooting
.