5-2
Instruction Manual
IM-106-4050 Original Issue
March 2006
OxyBalance Oxygen Display 
and Averaging System
PanelView Removal
NOTE
Make sure to record the wiring connections and locations before 
disconnecting any wiring. This will aid in the installation of the unit.
1. Disconnect the three power wires from the wire contacts (L1, L2N, and 
ground) of the PanelView Plus 600, see Figure 5-1
2. Remove the communications cord from the PanelView Plus 600.
3. Turn mounting levers to the unlocked position, and remove the 
mounting levers from the mounting lever slots. See Figure 5-3.
4. Carefully slide the PanelView Plus 600 unit through the front of the 
panel cutout.
Figure 5-1. PanelView System 
Wiring
Disconnect all power before installing or replacing components. Failure to disconnect power 
may result in electrical shock and/or damage to the equipment.
Power
Connection
38370010
Communications
Cable