Emerson Process Management NGA2000 HFID 用户手册

下载
页码 76
Instruction Manual 
748297-E 
June 2003 
4-14     Maintenance and Troubleshooting 
Rosemount Analytical Inc.    A Division of Emerson Process Management  
Model NGA2000 HFID
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
Figure 4-12. 
Electronics Assembly – Exploded View 
 
 
 
Printed Circuit Boards 
When replacing a circuit board, the follow-
ing procedure is recommended: 
1.  Per Section 4-2e on page 4-13, re-
move securing hardware from elec-
tronics assembly and lay on bench. 
2.  Remove securing hardware from 
printed circuit board to be replaced, 
do not disconnect cable(s). 
 
3.  One at a time, remove the wiring con-
nectors and attach to replacement 
board. 
Mount replacement board to electronics 
assembly. 
 
 
 
 
 
 
 
 
Safety Board
657499 
Sensor Board 
657060 
Computer Analysis Board 
659149 
Preamp Assembly 
656945 
Case Pressure Purge 
Switch  903690 
Power Supply Board 
655764 
Case Temperature Sensor 
656026