Emerson Process Management NGA2000 HFID 用户手册
Instruction Manual
748297-E
June 2003
Rosemount Analytical Inc. A Division of Emerson Process Management
Maintenance and Troubleshooting 4-17
Model NGA2000 HFID
Preamp Assembly
1. Per Section 4-2e on page 4-13, re-
move securing hardware from elec-
tronics assembly and lay on bench.
tronics assembly and lay on bench.
2. Disconnect and note location of ca-
bles.
3. Remove the two screws and washers
from the top bracket and slide the
preamp assembly out.
preamp assembly out.
4. Remove the lower bracket from the
preamp assembly and install on re-
placement preamp assembly.
placement preamp assembly.
5. Slide replacement preamp assembly
into top bracket and secure with
mounting hardware.
mounting hardware.
6. Re-connect
cables.
Figure 4-15.
Preamp Assembly Installation
Bottom Bracket
Top Bracket
Preamp Assembly
656945
656945