Emerson Process Management NGA2000 HFID 用户手册

下载
页码 76
Instruction Manual 
748297-E 
June 2003 
Rosemount Analytical Inc.    A Division of Emerson Process Management  
Maintenance  and Troubleshooting     4-17 
Model NGA2000 HFID 
 
Preamp Assembly 
1.  Per Section 4-2e on page 4-13, re-
move securing hardware from elec-
tronics assembly and lay on bench.  
2.  Disconnect and note location of ca-
bles. 
3.  Remove the two screws and washers 
from the top bracket and slide the 
preamp assembly out. 
4.  Remove the lower bracket from the 
preamp assembly and install on re-
placement preamp assembly. 
5.  Slide replacement preamp assembly 
into top bracket and secure with 
mounting hardware. 
6. Re-connect 
cables. 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
Figure 4-15. 
Preamp Assembly Installation 
 
Bottom Bracket 
Top Bracket 
Preamp Assembly 
656945