4-28     Maintenance and Troubleshooting 
Rosemount Analytical Inc.    A Division of Emerson Process Management  
Model NGA2000 HFID
Fuel In 2-Way Solenoid Valve 
1.  Disconnect wiring solenoid valve wir-
ing connector, note location. 
2.  Inside the analyzer module, discon-
nect the tube going to the connector 
on the “out” port of the solenoid valve. 
3.  On the rear of the analyzer module at 
the fuel in port: 
a. 
Disconnect the fuel in tube. 
b. 
Remove nuts and washers. 
4.  Remove solenoid valve from analyzer 
module 
5.  Remove the fittings from the solenoid 
valve and replace the Teflon pipe 
thread tape. 
6.  Verify that body of replacement sole-
noid valve is oriented as shown in 
Figure 4-22 on page 4-27.  If not, ro-
tate till wires are in-line with “out” port. 
7.  Install fittings into replacement sole-
noid valve, re-install in analyzer mod-
ule. 
Burner Air In Filter 
1.  Leaving the bulkhead fitting secured 
to the rear panel, remove the tubes, 
nuts and ferrules from the fitting. 
2.  Insert a clean, rigid piece of tube or 
rod (smaller than .25 inch diameter) 
into the bulkhead fitting to force out 
the filter disc. 
3.  Install the replacement filter in the 
same manner, through the rear of the 
bulkhead fitting. 
4. Re-connect 
tubes. 
 
 
Heated Bypass Sample Out and Heated 
Sample In Restrictors 
1.  On the outside of the rear panel, dis-
connect tube and remove nut. 
2.  Insert a small spade screwdriver into 
the bulkhead and remove the restric-
tor. 
3.  Install in reverse order. 
Regulated Air In Check Valve 
1.  Disconnect tube at elbow. 
2.  Remove check valve from female 
connector. 
3.  Remove elbow from check valve. 
4.  Add Teflon pipe thread tape to check 
valve threads. 
5.  Install elbow onto check valve. 
6.  Install check valve into female con-
nector, verifying orientation of elbow 
fitting as shown in Figure 4-22 on 
page 4-27.