Teledyne Percent Oxygen Analyzer 用户手册

下载
页码 51
Percent Oxygen Analyzer  
Operational Theory 
 
 
 Teledyne 
Analytical 
Instruments 
17
 
Probes of 1000 mm (40”) normally have sufficient length for any 
installation.  Customers requiring probes longer than 1500 mm (59”) are 
supplied with a flow guide tube that uses the flue velocity to pull flue 
gas through a filter at the sensing tip and exhaust it near the flue wall. 
2.2.1 Probe Configuration 
The probe can be configured in a variety of ways to suit the 
analytical environment. Figure 2-1 shows the most common 
configurable items of the probe. Not indicated is the actual 
thermocouple sensing element, however, see Table 2-1. You can choose: 
1.  Probe insertion length (from process end of mounting thread to 
probe sensing tip).  See Figure 2-2. 
2.   Mounting thread (process connection), BST or NPT (for size of 
thread refer to specifications).  
3.   Lagging extension length, if required. 
Table 2-1 indicates typical probe options. 
Table 2-1: Probe Options 
Probe Length 
Thermocouple 
Sheath 
Thread 
250 mm (10”)  
Type K max 900°C (1560°F) 
316 SS max 850°C (1560°F) 
BST or NPT 
350 mm (14”)  
 
Inconel 
 
500 mm (20”)  
 
 
750 mm (30”)  
 
 
1000 mm (40”) 
 
The Inconel option has all 
inconel wetted parts except for 
the ceramic sensor and viton 
‘o’ rings 
 
 
 
 
Figure 2-2: Model 9060L Probe