Maintenance 
Aside from normal cleaning and checking for leaks at the probe 
connection, the Model 9060L should not require any maintenance 
beyond replacement of a blown fuse. Routine maintenance includes 
occasional recalibration, as described in Chapter 4, Operation
5.1 Replacing the Fuse 
CAUTION: 
REMOVE POWER TO UNIT BEFORE REPLACING 
THE FUSE. 
 
 
When a fuse blows, check first to determine the cause, then replace 
the fuse using the following procedure: 
1.  Disconnect the AC power and place the power switch located on 
the rear panel in the O position. Remove the power cord from the 
receptacle. 
2.  The fuse receptacle is located in the power cord receptacle 
assembly in the upper left-hand corner of the rear panel. See 
Figure 5-1. 
 
Figure 5-1: AC Fuse Replacement