Drive Medical Design 18605 用户手册

下载
页码 16
set up & operation
7
BATTERY OPERATION: To ensure proper operation from internal battery, fully charge the  
battery for 1.5 hours as explained in the Battery Charging section. To operate the unit from an 
internal rechargeable battery, ensure that no external power sources are plugged into the DC 
power input on the side of the unit.
Once the power source is selected, simply turn the unit on using the power switch located on the 
side of the unit. The power indicator light (displayed on the front of the unit) will glow yellow to 
show which power source is being used and glow green when an external power is connected.
battery charging/ cleaning
IMPORTANT OPERATION AND BATTERY INSTRUCTIONS: 
Your 18605 portable suction machine is capable to run on AC or DC power. To ensure full func-
tion of this unit, the battery must be completely charged before the first use. Do not turn on the 
unit before the battery is completely charged. It takes approximately 1.5 hours to fully charge 
the unit. Charge the unit when battery levels are low to ensure proper operation.
WARNING
If the unit does not receive an external power source or the battery does not get recharged im-
mediately, the low battery indicator light will remain on and the performance of the unit will drop 
off rapidly. Switch to another power source immediately to avoid an interrupted suction 
procedure. 
Adjust the vacuum level from 0 to 600 mm Hg by turning the vacuum regulator knob located on 
the side of the unit (clockwise to increase vacuum and counter-clockwise to decrease vacuum). 
The desired level of vacuum can be set by using the gauge located on the front of the unit.  To 
accurately read the gauge, block the patient end of the hose or cap off the collection bottle and 
allow the gauge to reach a stable vacuum reading. 
NOTE
Gauge is for reference only. If the unit sustains a severe drop, accuracy of the gauge must be 
checked. 
Suction at the patient is automatically obstructed when liquid level reaches the float shut-off 
located on the underside of the collection bottle lid. 
NOTE
Always transport unit with vacuum regulator knob rotated fully clockwise in case unit is dropped. 
CAUTION 
When automatic float shut-off is activated, contents of the collection bottle should be emptied. 
Further suctioning could cause damage to the vacuum pump. 
CAUTION 
Should fluid be aspirated back into the unit, equipment provider servicing is necessary as pos-
sible vacuum pump damage may result.