Emerson 755A 用户手册

下载
页码 90
Instruction Manual
245364-V
May 2002
Rosemount Analytical Inc.    A Division of Emerson Process Management                                        Description and Specifications        1-5
Model 755A
c. Electrical
Supply Voltage and Frequency
Standard ................................ 115 VAC 
±10 VAC, 50/60 Hz
Optional ................................. 230 VAC 
±10 VAC, 50/60 Hz
Power Consumption
Maximum ............................... 300 watts
Nominal..................................  75 watts
Output
Standard ................................ Field selectable voltage output of 0 to 10mV,  0 to 100mV,  0 to 1V,
or  0 to 5VDC
Optional .................................  Isolated current output of 0 to 20mA  or  4 to 20mA (with Current
Output Board)
Alarm Option.................................. High-Low Alarm
Contact Ratings ..................... 5 amperes, 240V AC, resistive load
5 amperes, 120V AC, resistive load
5 amperes, 28V DC, resistive load
Setpoint ......................................... Adjustable from 1% to 20% of fullscale
Deadband ...................................... Adjustable from 1% to 20% of fullscale  (Factory set at 10% of
fullscale)
d. Physical
Mounting
Standard ................................ Panel mount
Optional .................................  Surface or stanchion mount accessory available
Enclosure Classification ................ Meets requirements for NEMA 3R
Air Purge Option
1
................... NFPA 496 (1989) Type Z purge
Weight ........................................... Approximately 32.5 lbs (14.74 Kg)
Dimensions.................................... Height:  13.5 (343 mm)
Width:  11.5 (294 mm)
Depth:  7.12 (181 mm)
                                                     
1
 When installed with user supplied components, meets requirements for Class I, Division 2 locations per National Electrical
Code (ANSI/NFPA 70) for analyzers sampling nonflammable gases.  Analyzers sampling flammable gases must be pro-
tected by a continuous dilution purge system in accordance with Standard ANSI/NFPA 496-1986, Chapter 8.  Consult factory
for recommendations.