Spectral Absorption of Hb and HbO
2
SpO
2
 100 (% SpO
2
)
C(HbO
2
C(HbO
2
)+C(Hb)
Absorption coef
ficient
Wavelength
Red light
Infrared light
Measurement Principle
This instrument measures the oxygen saturation (SpO
2
) in arte-
rial blood and pulse rate continuously in a non-invasive method 
by applying optical principles.
With this instrument, SpO
2
 is defined by the following equation.
where
C (Hb): 
Concentration of reduced hemoglobin
C (HbO
2
):  Concentration of oxyhemoglobin
The light-absorption characteristics of reduced hemoglobin (Hb) 
are very different from those of oxyhemoglobin.  This instrument  
measures the changes in the absorption of red and infrared 
lights passing through the tissue to determine the SpO
2
 of the 
blood.  Thus, this method is free from the effects of skin color, 
muscles, bones, and veins.