Troubleshooting
5-10
Monitoring SpO
2
Troubleshooting
The table below lists system messages that you may encounter when monitor-
ing SpO
2.
Table 5-2  System Messages
Problem or Message
Possible Cause
Corrective Action
6S21RQ3XOVDWLOH
z
Pulse absent or too 
weak to be detected.
z
Check the sensor is 
applied properly.
z
Make sure the sensor 
site has a pulse. 
z
Relocate the sensor to 
a site with improved 
circulation.
z
Try another sensor 
type.
6S2/RZ6LJQDO
z
SpO
2
 signal is too low 
to give an accurate read-
ing.
z
Check the sensor is 
applied properly.
z
Try another sensor 
type.
6S21RLV\6LJQDO
z
Excessive patient move-
ment, electrical interfer-
ence, or optical 
interference.
z
Minimize patient 
motion or apply sen-
sor to site with less 
movement.
z
Secure the sensor 
cable loosely to the 
patient.  
z
Reduce sources of 
electrical or optical 
interference. 
6S2/LJKW,QWHUI
z
The level of ambient 
light is so high that the 
sensor cannot obtain an 
SpO
2
 reading.
z
Sensor or cable is dam-
aged.
z
Cover sensor with an 
opaque material.
z
Check sensor for dam-
age; try another sen-
sor.