Defibrillating
6-4
Defibrillating in Manual Mode
Defibrillating
The defibrillation process is as follows:
1. Select Energy
The default energy setting is 150 joules. To change the energy setting, 
press 
 repeatedly to cycle through the energy level choices. Stop 
when the desired energy is displayed. The choices are 5, 10, 25, 50, 70, 
100, 150, or 200 joules.
Figure 6-3  Energy Select Display
2. Charge
Press . 
As the defibrillator charges, the current charge is displayed above the 
shock counter. A charging tone beeps until the desired energy level is 
reached, at which point you’ll hear a continuous charge tone. 
(1(5*<
6(/(&7
ENERGY
SELECT
HR ALARM
LEAD
SELECT
CHARGE
Shocks:0
00:00:8
78
Pads
SpO2
150 J
&+$5*(