Philips 861304 用户手册

下载
页码 64
38
Ph
ilip
s M
e
di
ca
l Sy
st
e
m
s
HEARTSTART FRx 861304 OWNER’S MANUAL
AP
P
E
ND
IC
ES
Accessories specifications
HeartStart SMART Pads II 989803139261 
non-shockable rhythm — all
other non-shockable
rhythms
c
450
specificity >95%
(88%)
a. From Philips Medical Systems Heartstream ECG rhythm databases.
b. American Heart Association (AHA) AED Task Force, Subcommittee on AED Safety & Efficacy. Automatic External Defibrillators for Public 
Access Use: Recommendations for Specifying and Reporting Arrhythmia Analysis Algorithm Performance, Incorporation of New 
Waveforms, and Enhancing Safety. Circulation 1997;95:1677-1682.
c. Supraventricular tachycardia (SVT) is specifically included in the non-shockable rhythm class, in accordance with AHA recommendations
b
 
and the AAMI standard DF80.
rhythm class
ECG test
sample
a
 size
meets AHA recommendations
b
 for adult defibrillation
observed 90% 
one-sided
performance
lower
confidence limit
category
specifications
pads for defibrillation,
pacing, monitoring,
cardioversion
Disposable, adhesive pads with a nominal active surface area of 80 cm
2
 each, 
provided in a disposable plastic case, and an integrated 48 inch (121.9 cm), 
typical, cable. Pads in case are designed to fit into carry cases.
SMART Pads II compatibility
defibrillator model
adult
patient use
infant/child patient use
FRx*
FR2/FR2+
FR/ForeRunner
MRx/XL/XLT/4000
HS1/OnSite/Home
competitive adapters
yes
yes
yes
yes
no; use M5071A
yes
yes
no, use M3870A
no
manual mode only
no; use M5072A
manual mode only
* Pre-connectable to FRx defibrillator only.
pads shelf life
Pads package is labeled with a use-by date of at least two years from date of 
manufacture.