Disconnect the main air tubing from the swivel.
The R
ES
M
ED
 H
OSPITAL
 N
ASAL
 M
ASK
 is for single patient, multi-use. The mask may be 
cleaned using a 70% v/v isopropyl alcohol swab or by wiping with a damp, soapy cloth. If 
using a damp cloth, use mild soap or diluted dishwashing detergent. Allow to air dry out of 
direct sunlight.
Note: Do not disassemble the mask.
CAUTIONS
Do not use solutions containing bleach, chlorine, aromatics, moisturizers, 
antibacterial agents, or scented oils to clean any part of the system or air tubing. 
These solutions may cause damage and reduce the life of the product.
Do not expose any part of the system or tubing to direct sunlight as it may 
deteriorate.
If any visible deterioration of a component is apparent (cracking, crazing, tears, etc), 
replace the mask.
Check the vent holes for blockage. Ensure the vent holes are cleaned regularly.
T
ECHNICAL
 S
PECIFICATIONS
P
RESSURE
 / F
LOW
 C
URVE
Note: As a result of manufacturing variations, the vent flow may vary from the values listed in 
the following table. 
F
LOW
 G
ENERATOR
 S
ETTING
Select “Standard” as the mask option, when using the R
ES
M
ED
 H
OSPITAL
 
N
ASAL
 M
ASK
 
with ResMed flow generators that have mask setting options.
D
EAD
 S
PACE
 I
NFORMATION
Explanation: Dead space is the empty volume of the mask up to the swivel.
The dead space of the mask varies according to cushion sizes, but is less than 104 mL.
E
NVIRONMENTAL
 C
ONDITIONS
Operating Temperature: 41°F (5°C) to 104°F (40°C)
Operating Humidity: 15–95% relative humidity
Pressure
(cm H
2
O)
Flow
(L/min)
4
19
8
28
12
34
16
40
20
45
0.0
5.0
10.0
15.0
20.0
25.0
30.0
35.0
40.0
45.0
50.0
4
6
8
10
12
14
16
18
20
Pressure (cm H
2
0)
F
lo
w Rate (L
/m
in
)