Disinfecting/Sterilizing the Mask or Nasal 
Pillows System between Patients 
This Disinfection/Sterilization Guide is intended for multipatient use of the 
mask/nasal pillows
*
 system in a sleep lab, clinic or hospital environment. If you 
use the mask as a single user in the home, refer to the User Guide for cleaning 
instructions.
The following are the recommended and validated procedures for cleaning, 
disinfection and sterilization of the mask. We advise that each healthcare 
facility consult its own procedures before carrying out the following 
instructions. 
For details on the mask refer to the User Guide. Parts that cannot be 
disinfected/sterilized are listed in the table on page 5.
 WARNINGS AND CAUTIONS
ResMed cannot give any assurance that deviations from the 
procedures listed below, and their effect on the performance of the 
product, will be acceptable.
The mask components should not be subjected to autoclave or 
ethylene-oxide gas sterilization.
When using detergents, disinfectants or sterilization agents, always 
follow the manufacturer’s instructions.
Notes: 
The headgear does not require disinfection or sterilization. However, it is 
necessary to thoroughly wash the headgear before using between patients. 
The cleaning procedure is detailed in the User Guide.
The mask must be disassembled and cleaned prior to disinfection or 
sterilization. The mask must also be inspected after disinfection/sterilization. 
See the section "Inspection after Disinfection/Sterilization" for inspection 
details.
* References to mask/s in this guide refer to ResMed masks and nasal pillows system.