Invacare IRC5PO2 用户手册

下载
页码 88
SECTION 12—HEAT EXCHANGER ASSEMBLY
51
SECTION 12—HEAT EXCHANGER 
ASSEMBLY
Replacing Heat Exchanger Assembly
 DANGER
To prevent electrical shock, ALWAYS disconnect concentrator from electrical outlet 
before servicing.
NOTE: For this procedure, refer to FIGURE 12.1 on page 52.
NOTE: The heat exchanger assembly contains three components: heat exchanger, nine inch vinyl 
tube and connecting component.
1. Unplug unit.
CAUTION
Use care not to deform heat exchanger when installing, removing or cleaning.
3. Disconnect brass nut that secures the heat exchanger assembly to the compressor 
(Detail “A”).
4. On the backside of the Perfecto2, loosen the clamp that secures the vinyl tube to the 
manifold assembly and remove the vinyl tube from the manifold assembly. (Detail 
“B”).
5. Remove heat exchanger assembly by pulling vinyl tube through sound box.
6. Reverse STEPS 3‐5 to install new heat exchanger.
7. Run unit and check for leaks. Refer to Leak Test on page 76.