SECTION 15—P.C. BOARD
58
SECTION 15—P.C. BOARD 
Replacing P.C. Board
 DANGER
To prevent electrical shock, ALWAYS disconnect concentrator from electrical outlet 
before servicing.
CAUTION
Follow these pre‐cautions to prevent damage to the P.C. boards:
Before handling any P.C. boards, you need to be properly grounded to prevent static 
damage to the components of the board. A Static Cuff MUST be worn and properly 
grounded using an alligator clip. Electrical conduit or a water pipe is normally sufficient 
when a known good ground is not available. Care should be taken to ensure that the 
alligator clip contacts with bare metal surface.
When removing quick disconnects terminals, DO NOT pull on wire itself as damage to 
the connection may occur. Hold down the P.C. board with one hand and use an upward 
force with a slight rocking motion to remove the terminals.
Before installing any P.C. boards, ensure that all insulators are in place.
NOTE: For this procedure, refer to FIGURE 15.1 on page 60.
1. Unplug unit.
4. Do the following:
A. Disconnect wiring harnesses from PC board.
 WARNING
DO NOT remove oxygen sensor tubing from P.C. board. Remove from regulator ONLY. 
DO NOT remove the pressure sensor tubing from P.C. board. Remove from top of 
product tank only. Damage to oxygen sensor assembly or pressure sensor could occur. A 
faulty sensor may result in improper oxygen delivery and harm to the user.
B. Disconnect pressure sensor tubing from product tank, not P.C. Board.
C. For concentrators that have O2 sensor, disconnect the oxygen sensor tubing from 
5. Remove the two mounting screws on top corners of P.C. board.
6. Position new P.C. board.