SECTION 24—ALARM TEST
82
FIGURE 24.2   High Pressure Test
Oxygen Sensor 
NOTE: For this procedure, refer to FIGURE 24.3 on page 83.
Alarm sequence occurs when concentration levels fall below a preset value, typically 73% 
± 3% or 85% ± 2%. This failure indicates a catastrophic failure of any component 
mechanical or electrical. 
1. Turn unit on ( ). Set output flow at 5 L/min.
2. Monitor the O
2
 level.
3. With the O
2
 level greater than 85% after five minutes, the GREEN panel indicator light 
will remain illuminated.