ResMed S8 Escape II 用户手册

下载
页码 80
6
How to Change the Settings on your S8 Escape II
You can change settings (such as Ramp Time) on the S8 Escape II by using the 
keypad and LCD. When the S8 Escape II is in standby mode you can display a 
series of screens (a menu) on the LCD screen in order to view and change the 
settings for a particular function. 
Use the menu screens to change the ramp time and language. You can also view 
details about the usage and software version of your S8 Escape II. 
How to change the ramp time
During ramp time, the pressure increases from a low pressure to the prescribed 
treatment pressure. You can set the ramp time in 5-minute intervals, from 0 
minutes (OFF) up to a maximum ramp time set by your clinician. The standby 
screen displays the current ramp time in the right corner. 
1. Press 
 or 
 until the desired ramp time appears.
2. Once a ramp time is selected, begin therapy by pressing 
.
When ramping begins the LCD will display the word "RAMPING", with the 
current pressure indicated as dashes. Each dash is approximately equivalent 
to 2 cm H
2
O. For example, six dashes represents approximately 12 cm H
2
pressure. 
How to change the EPR level 
EPR can be used to help ease any discomfort you may experience when 
breathing out during treatment. If your clinician has allowed access you can 
change the EPR level. There are four settings: OFF; 1 (lowest EPR); 2; and 3 
(highest EPR).
S8 CLASSIC
S8 CLASSIC
S8 ESCAPE II
PATIENT MENU
RAMP 
10min
Menu 
 EPR LEVEL: 1
 Change 
 Exit
USED HRS: 
960
 
 Exit
USAGE: 
120/180
 
 Exit
LANG:  
ENGLISH
Change 
 Exit
SW: 
SX123456789
 
 Exit
Only displayed if 
clinician enabled
338518 American Multi.book  Page 6  Tuesday, February 5, 2008  2:07 PM