Clinical Use Only
4
Option 3: STERRAD Sterilization System
Note: The Ultra Mirage™ II Nasal Mask is not validated for the STERRAD 
Sterilization System.
Pre-sterilization
1 Disassemble the mask.
2 Clean the individual parts of the mask with a soft bristle brush for one 
minute while soaking them in detergent (eg, Alconox). Pay particular 
attention to all crevices and cavities.
3 Rinse the components twice by agitating them vigorously in drinking quality 
water and allow them to air dry out of direct sunlight.
ResMed has tested the detergent Alconox (diluted at 1%).
Sterilization
Sterilize the sterilizable mask components as described in the manufacturer’s 
instructions for the STERRAD Sterilization System (eg, Sterrad 100S or 
Sterrad NX).
Testing by ResMed has shown that the mask components can withstand 15 
cycles of the STERRAD 100S Sterilization System, when using the tray 
method.
Note: The use of pouches is not recommended.
Inspection after Disinfection/Sterilization 
Perform a visual inspection of each mask component. If any visible 
deterioration of a mask component is apparent (cracking, crazing, tears etc), 
the mask component should be discarded and replaced. Slight discoloration of 
the silicone components may occur and is acceptable.
Note: If a healthcare facility requires an additional disinfection or sterilization 
process after reassembly, the number of validated cycles is halved.
Parts not Validated for Multipatient Use
Some parts of the masks have not been validated for multipatient use. These 
parts need to be replaced with new parts when using between patients. 
Please check the Components Card on the website www.resmed.com for a list 
of available replacement parts for each mask system.