Troubleshooting
Excessive leak.
Adjust position of mask and headgear.
Replace caps if missing from ports on mask. 
Air tubing may not be connected properly. Connect 
firmly at both ends.
Air tubing may be kinked or punctured. Straighten or 
replace.
Displays error message: 
SYSTEM ERRxxxx Call Service!
 or 
Exxxx
 (where 
xxxx defines an error)
Component failure.
Record error number and call service.
Displays error message: 
Check tube!! Key if done
Air tubing is loose or blocked.
Check that the air tubing is connected securely to your 
mask and the air outlet to the front of the S8 Series II. 
When you have checked the air tubing, press the Start/
Stop key to restart the device. If this does not clear the 
message, disconnect the power cord and then 
reconnect it to restart the device.
The following message is displayed on the LCD after you try to update settings 
or copy data to the Data Card: 
Card Error Remove Card
Data Card is not inserted correctly.
Check that the Data Card is inserted into the Data Card 
module with
 
the arrow facing up, while the S8 Series II 
is switched on and not delivering therapy (standby 
mode).
The Data Card is only partly 
inserted.
Ensure that the Data Card is inserted into the Data Card 
slot as far as it will go.
You may have removed the Data 
Card before settings were copied to 
the S8 II.
Reinsert the Data Card and wait for the 
Settings
 
Success Remove Card
 message to appear on the 
LCD.
The card is not a ResScan Data 
Card.
Only use a ResScan Data Card.
The following message is displayed on the LCD after you try to update settings 
or copy data to the Data Card: 
Settings Error Remove Card
There is a data error on the Data 
Card.
Contact your clinician immediately.
The following message is NOT displayed on the LCD after you try to update the 
settings using the Data Card: 
Settings Success Remove Card
The settings were not updated.
Contact your clinician immediately.
Problem/Possible Cause
Solution