5. At step 3, pressing the  PRINT  key to weigh the 
zero-point value. Do not touch the pan during 
weighing. 
 
 
 
 
 
 
 
 
 
 
 
 
6. Place the calibration weight with the same value as 
displayed on the pan. Press the  PRINT  key to 
weigh it. Do not touch the pan during weighing. 
 
 
 
 
 
 
 
 
7.  end     appears. 
Remove the weight from the pan. 
When the GLP output (function setting “info 1” or 
“info 2”) is selected,  glp     is displayed. The 
calibration report is output and  end     appears 
again. 
Press the CAL switch or  UNITS  key to return to 
the weighing mode. 
 
 
 
 
 
 
 
 
‰
 Note 
  The value set in step 4 is stored in memory even after the power is 
switched off. 
 
  If the balance is to be moved to other places, set the gravity 
acceleration value of the area where the calibration using a weight is 
to be done, and calibrate the balance according to the procedure 
above. See the next section to set the value. 
 
 
 
  5000.0 
 
0
end
000
 
With nothing 
on the pan 
To store
To store
Calibration
weight 
  Cal 0 
  5000.0 
  glp 
GLP output
 
0
end
000
 
Returns to   
the weighing mode