Change the function table   
Selecting a way to set the density of a liquid 
Select the liquid density input method from the function table below. The function table 
is available only when the density measurement mode is selected. 
Class Item 
Param- 
eter 
Description 
Š 
Water temperature 
func 
ldin 
Liquid density input 
  1 
Liquid density 
The way to input 
liquid density. 
 
Š
 Factory setting 
 
Setting the density of a liquid   
1. Press the    UNITS    key to select    SG   . 
 
 
2. Press and hold the    UNITS    key to display liquid density input mode. 
ldin = 0: Water temperature 
This shows 25 °C. 
 
ldin = 1: Liquid density 
This shows ρ = 1.0000 (g/cm
3
). 
 
 
3. Using the   RE-ZERO   (to increment the value) and   SAMPLE   key (to shift the 
selected digit), set the value and press the    PRINT    key to store. 
‰
 To cancel the setting procedure and return to the density measuring mode, press 
the  
  UNITS    key. The input value is not stored. 
‰
 The relation between the water temperature and density is shown below. 
 
°C 
+0 +1 +2 +3 +4 +5 +6 +7 +8 +9 
0.99984 0.99990 0.99994 0.99996 0.99997 0.99996 0.99994 0.99990 0.99985 0.99978
10 
0.99970 0.99961 0.99949 0.99938 0.99924 0.99910 0.99894 0.99877 0.99860 0.99841
20 
0.99820 0.99799 0.99777 0.99754 0.99730 0.99704 0.99678 0.99651 0.99623 0.99594
30 
0.99565 0.99534 0.99503 0.99470 0.99437 0.99403 0.99368 0.99333 0.99297 0.99259
40 
0.99222 0.99183 0.99144 0.99104 0.99063 0.99021 0.98979 0.98936 0.98893 0.98849
50 
0.98804 0.98758 0.98712 0.98665 0.98618 0.98570 0.98521 0.98471 0.98422 0.98371
60 
0.98320 0.98268 0.98216 0.98163 0.98110 0.98055 0.98001 0.97946 0.97890 0.97834
70 
0.97777 0.97720 0.97662 0.97603 0.97544 0.97485 0.97425 0.97364 0.97303 0.97242
80 
0.97180 0.97117 0.97054 0.96991 0.96927 0.96862 0.96797 0.96731 0.96665 0.96600
90 
0.96532 0.96465 0.96397 0.96328 0.96259 0.96190 0.96120 0.96050 0.95979 0.95906
 
 
Example of density measurement
 
Selecting the SG measurement mode 
1. Press the    UNITS    key to select    SG   . 
  
 
 
 
(The weight unit is “g”.) 
‰
 The weighing unit is “g”. 
‰
 The display shows that 
 HI 
  blinks and the balance   
is measuring weight in air. 
‰
 When the display does not show zero, press the  RE-ZERO  key to set the 
display to zero.   
 t
---
25 
SG 
 d1.0000 
SG 
SG indicator
    0.00 
SG 
HI 
SG indicator
    0.00 
SG 
HI 
Measuring 
weight in air.