Tanita 4521 用户手册

下载
页码 9
6
TROUBLESHOOTING
The following guide will help you to solve 
common problems:
TRANSPORTING AND STORING
See Fig. 5 for visual assistance.
1. Unplug the adapter 
from the outlet, then
unscrew it from the 
electronic control 
box and set the 
adapter aside.
2. Slide the adjustable 
rail to its widest setting.
3. Loosen the locking screw 
and slide it off the black tabs.
4. Slowly lift the scale by the handle.
Avoid letting the rails bang together.
5. Wrap an elastic cord (64010) around each end of 
the rails and hook the cord ends together.
6. Unscrew the scale cable from the control box and
firmly place the control panel box on the velcro tape 
of the center column. Wrap the cable around the 
center column above the control panel box and secure.
7. The scale is now ready for transportation or storage.
8. Remember to keep the adapter with the scale.
9. For additional convenience and protection of your
portable wheelchair scale, a scale storage bag is
available from your Tanita representative.
KILO
S
ON
OFF
Pounds
Kilos
www.ta
nita.co
m
Push “
ON” b
utton
and w
ait un
til zero
appea
rs befo
re
steppin
g on s
cale
WEIGHT
POUND
S
ZERO
Fig. 5
PROBLEM
Incorrect patient 
weight readings
Patient weight readings
consistently too high 
or too low
Display flashes “Lo-b”
SOLUTION
A. Platform is too low and hitting platform feet or
carpet. Increase platform height. See “Leveling
the Platform” on page 2.
B. Person, other than person being weighed, 
is touching or leaning on scale handlebar
C. Weight of wheelchair or chair is not taken into
account. See “Operating Instructions” on page 4.
D. Platform is not level. See “Leveling the Platform”
on page 2.
Check scale calibration and calibrate if necessary.
See “Calibration” on page 5.
Recharge battery. See “Charging the Battery” on
page 3.