Monarch 315 用户手册

下载
页码 6
MODEL 315  HUMIDITY TEMPERATURE METER 
 
3.0 Symbol Definition and Button Locations: 
 
 
 
 
 
 
 
 
Feature Locations: 
1) Dust 
mask 
2) LCD 
display 
3) ON/OFF 
button 
4)  MAX MIN button 
5) HOLD 
button 
6) RELATIVE 
button 
7)  °C, °F control button 
8) Battery 
cover 
 
HUMIDITY:
TEMP:
4
6
7
5
3
2
F
140
4 F
0
%RH
100
MIN
MAX
¢XF
C
20
¢XC / 
C
60
REL
HOLD
8
1
HUMIDITY
OPEN
9V  BAT
NEDA 1604  6F22  006P
TERY
Auto power off
indication
Temperature display
Unit indication
Data Hold indication
Minmum indication
Low battery indication
Maximum indication
Relative indication
Humidity display
MODEL 315  HUMIDITY TEMPERATURE METER 
 
 
5.2 Humidity 
Calibration 
1.  Turn the unit off. Press and hold the HOLD and °C/°F buttons. While 
these two buttons are pressed, turn the power on. Release all buttons 
and all the segments on the display will blink. 
2.  Within 3 seconds of releasing these buttons, press the 
UREL button to 
enter the calibration mode, or the instrument will revert to normal 
operating mode. 
3.  Upon entering the calibration mode, ‘CAL1’ will appear on the second 
display.  
4.  Place the probe in a stable 32.8% RH environment at 25°C and allow 
the system to stabilize for 30 minutes. Press the MAX MIN button to 
initiate calibration. A single long beep will indicate calibration is 
complete, and the instrument will advance into the ‘CAL2’ mode, as 
indicated on the second display. 
5.  Insert the probe in a stable 75.3% RH environment at 25°C and again 
allow the system to stabilize for 30 minutes. Press the MAX MIN button 
to initiate calibration. A single long beep will indicate calibration is 
complete, and the instrument will revert to normal operating mode. 
5.3  Restoring Factory Calibration 
1.  To restore factory default calibration values, turn the instrument off. 
Press the HOLD and °C/°F buttons. While these two buttons are 
pressed, turn the power on. Release all buttons and all the segments 
on the display will blink. 
2.  Within 3 seconds of releasing these buttons, again press the HOLD 
button to restore factory calibration values. A single long beep will 
indicate restoration is complete.