Clarity - INT9 A/D converter 
 
Using the INT9 card 
17 
Select Detector 
Selects the detector whose parameters will be 
set. If there is only one detector (channel) 
configured on the Clarity Instrument, this 
control will not be displayed. 
Range 
Selects the input voltage range (in mV). The 
selected range corresponds to the maximum 
output voltage of the detector that the A/D 
converter can still process.  
Available ranges for INT9 card are 
156
1250
 
and 
12 123 mV
.  
Sample Rate 
Depending on 
Supply Frequency
 filter selected 
in the 
System Configuration - INT9 A/D Card Setup
 
dialog, the available sampling rates are 
6.3
12.5
25
50
100
200
 and 
400
 samples per 
second (for 50Hz mains frequency filter) or 
7.5
15
30
60
120
240
 and 
480
 samples per 
second (for 60Hz mains frequency filter).  
Higher sampling rate allows for the 
measurement of narrower peaks, but it also 
means a larger amount of data which will affect 
the size of the resulting chromatogram and the 
speed of its processing. 
Sample rate sufficient for successful peak 
detection is about 20 samples per narrowest 
peak. 
A higher sampling rate also increases the 
amount of noise.  
Note: 
Range
 and 
Sample Rate
 are set individually for 
each detector, as opposed to 
Mains Frequency
 set 
for the INT9 A/D card as a whole.