Clarity INT9 用户手册

下载
页码 35
Clarity - INT9 A/D converter 
 Troubleshooting 
29 
 
•  In the Models list select the 
DataApex 
Acquisition device, Model 9.00
 and press the 
Next
 button. 
•  In the remaining dialogs press the 
Next
 
button. 
•  Restart the PC. 
•  Re-check the driver in the 
Device Manager
 as 
desribed in chapter 4.3.3
4.4  Data Acquisition – non-functional 
A gray 
 icon, label “Disabled”, and Data Acquisition command not functional 
Other manifestations of this error are also: 
Method Setup - Acquisition
 tab missing, 
Method – 
Acquisition
 command non-functional, 
Run
Stop
Abort
 commands non-functional in the 
windows 
Single Analysis
 and 
Sequence
).  
Possible causes are: 
a)  Detector not allocated to instrument: 
Open the 
System Configuration
 dialog from the 
main 
Clarity
 window using the 
System - 
Configuration
 command and check the tab of 
the corresponding instrument - 
Instrument x
. If it 
has no allocated detectors, add them. 
Solution: 
In the left-hand list 
Setup Control Modules
 
select the correct detector that is connected to 
the A/D card you are using and link it to the 
corresponding instrument on the right.