General Troubleshooting & Repair 
Teledyne API T703 Calibrator Operation Manual 
212  
 
11.5.2. CHECKING MEASURE / REFERENCE VALVE 
To check the function of the photometer’s measure / reference valve: 
1.  Set the calibrator’s front panel display to show the PHOTO REFERENCE test function (see Section 6.1). 
photometer.  
 Press XZRO and allow the calibrator to stabilize. 
3.  Before completing the calibration by pressing the ZERO key, note of the displayed value. 
4.  Press the final Zero key, then press “NO” when asked, “ARE YOU SURE”. 
5.  Follow the instruction in Section 8.1.3.2 for performing a span point calibration of the photometer.  
 Press XSPN and allow the calibrator to stabilize. 
6.  Before completing the calibration by pressing the SPAN key, note of the displayed value of PHOTO 
REF
  If the O
3
 REF value has decreased by more than 2 mV from its value with Zero-gas, then there is a 
"cross-port" leak in the m/r valve.  
7.  Press the final Zero key then press “NO” when asked, “ARE YOU SURE”. 
 