6.9.2.2. Automatic TEST CHANNEL Calibration
To calibrate the outputs as a group with the
AOUTS CALIBRATION command, press:
NOTE
Before performing this procedure, make sure that the AUTO CAL feature is turned OFF for CONC_OUT_1
and CONC_OUT_2,
Make sure that the AUTO CAL feature is turned ON for the TEST CHANNEL (See Section 6.9.2.1)
From the
AIO CONFIGURATION SUBMENU
(See Section 6.9.1.1.)
Analyzer
automatically
calibrates all
channels for which
AUTO-CAL is turned
ON
If any of the channels have not
been calibrated ot if at least one
channel has AUTO-CAL turned
OFF, this message will read NO.
DIAG AIO
NOT AUTO CAL. CONC_OUT_1
DIAG AIO
NOT AUTO CAL. CONC_OUT_2
DIAG AIO
AUTO CALIBRATING TEST_OUTPUT
This message
appears when
AUTO-CAL is
appears when
AUTO-CAL is
Turned OFF for
a channel
DIAG AIO AOUTS CALIBRATED: YES
SET> CAL
EXIT
DIAG AIO AOUTS CALIBRATED: NO
SET>
CAL
EXIT
DIAG
ANALOG I/O CONFIGURATION
PREV NEXT
ENTR
EXIT
NOTE:
Manual calibration sho
tputs must be closely
uld be used for the 0.1V range or in cases where the ou
matched to the characteristics of the recording device.