TELEDYNE API 
Operating the M703E Calibrator 
M703E Calibrator Operator’s Manual 
70 
 
05744 Rev B 
6.9.2.2. Automatic TEST CHANNEL Calibration 
To calibrate the outputs as a group with the 
AOUTS CALIBRATION command, press: 
NOTE 
Before performing this procedure, make sure that the AUTO CAL feature is turned OFF for CONC_OUT_1 
and CONC_OUT_2, 
Make sure that the AUTO CAL feature is turned ON for the TEST CHANNEL (See Section 6.9.2.1) 
 
From the
AIO CONFIGURATION SUBMENU
(See Section 6.9.1.1.)
Analyzer 
automatically 
calibrates all 
channels for which 
AUTO-CAL is turned 
ON
If any of the channels have not 
been calibrated ot if at least one 
channel has AUTO-CAL turned 
OFF, this message will read NO.
DIAG AIO
NOT AUTO CAL. CONC_OUT_1
DIAG AIO
NOT AUTO CAL. CONC_OUT_2
DIAG AIO
AUTO CALIBRATING TEST_OUTPUT
This message 
appears when
AUTO-CAL is 
Turned OFF for 
a channel
DIAG AIO           AOUTS CALIBRATED: YES
SET>   CAL
EXIT
DIAG AIO           AOUTS CALIBRATED: NO
SET>
CAL
EXIT
DIAG              
ANALOG I/O CONFIGURATION
PREV   NEXT  
 ENTR
EXIT
 
 
 
NOTE: 
Manual calibration sho
tputs must be closely 
uld be used for the 0.1V range or in cases where the ou
matched to the characteristics of the recording device.