Data Acquisition System (DAS) and APICOM 
Teledyne API - T100 UV Fluorescence SO2 Analyzer  
160 
 
7.2.2. EDITING DAS DATA CHANNELS 
Although DAS configuration is most conveniently done through the APICOM remote 
control program (refer to Section 6.12.2.8), the following illustrations shows how to edit 
DAS channels using the analyzer’s front panel control buttons. 
 
 
Main Data Acquisition Menu 
Edit Data Channel Menu
SAMPLE          RANGE = 500.000 PPB        SO2 =XXX.X     
 
 < TST  TST >  CAL  
 
 
SETUP
 
 
 
 
SETUP X.X     
 0) CONC1:   ATIMER,  2,  4032, RS-232
 
 
PREV  NEXT
    
        INS    DEL   EDIT     PRNT    EXIT
 
SETUP X.X 
 
     
PRIMARY SETUP MENU
 
 
 CFG   
DAS
  
RNGE
  
PASS  CLK  MORE 
EXIT
 
SAMPLE          
ENTER SETUP PASS : 818
 
        
 8        1        8 
 
                      
ENTR
  
 EXIT
 
 
EXIT
 will return to  the 
previous SAMPLE 
display. 
SETUP X.X       
     
DATA ACQUISITION
 
 
 
 VIEW
 
 
 
 
 
 
EDIT
 
 
 
 
 
EXIT
 
Moves the 
display up & 
down the list of 
Data Channels 
Inserts
 a new Data 
Channel into the list 
BEFORE the Channel 
currently being displayed  
Deletes
 the Data 
Channel currently 
being displayed 
Exports the 
configuration of all 
data channels to 
RS-232 interface. 
Exits
 to the Main 
Data Acquisition 
Menu 
 
 
 
SETUP X.X      
NAME:CONC1
 
<SET SET>
 
 EDIT  PRNT       
 
EXIT
 
Moves the display 
between the 
PROPERTIES for this 
data channel. 
Reports the configuration of current 
data channels to the RS-232 ports. 
Exits
 returns to the 
previous Menu 
Allows to edit the channel name, see next menu button. 
 
 
Figure 7-3: DAS – Editing DAS Data Channels 