Teledyne API - T100 UV Fluorescence SO2 Analyzer  
Principles of Operation 
 
 
307 
conditioning circuitry on the PMT preamplifier board. Refer to Section 5.9.5 fo
instructions on performing this test. 
The optical test (OTEST) feature causes an LED inside the PMT cold block to create a 
light signal that can be measured with the PMT. If zero air is supplied to the analyzer, 
the entire measurement capability of the sensor module can be tested including the PMT 
and the current to voltage conversion circuit on the PMT preamplifier board. Refer to 
Section 5.9.4 for instructions on performing this test. 
13.5.6. PNEUMATIC SENSOR BOARD 
The flow and pressure sensors of the T100 are located on a printed circuit assembly just 
behind the PMT sensor. Refer to Section 12.6.15 on how to test this assembly. The 
signals of this board are supplied to the motherboard for further signal processing. All 
sensors are linearized in the firmware and can be span calibrated from the front panel. 
Refer to Section 5.4.3.2 for instructions on performing this test. 
13.5.7. RELAY BOARD 
The relay board is the central switching unit of the analyzer. It contains power relays, 
status LEDs for all heated zones and valves, as well as valve drivers, thermocouple 
amplifiers, power distribution connectors and the two switching power supplies of the 
analyzer. The relay board communicates with the motherboard over the I
2
C bus and is 
the main board for trouble-shooting power problems of any kind. 
13.5.7.1. HEATER CONTROL 
The T100 uses a variety of heaters for its individual components. All heaters are AC 
powered and can be configured for 100/120 VAC or 220/230VAC at 50-60 Hz.  
The two sample chamber heaters are electronically connected in parallel for analyzers at 
100/120 VAC line power and in series for units configured for 220/230 VAC. One 
configuration plug on the relay board determines the power configuration for the entire 
analyzer. 
On units with IZS options installed, an additional set of AC heaters is attached to the 
IZS permeation tube. Some special T100 models may have other, non-standard heating 
zones installed, such as a dilution manifold. 
13.5.7.2. VALVE CONTROL 
The relay board also hosts two valve driver chips, each of which can drive up four 
valves. In its basic configuration the T100 requires no valve control to operate. 
However, on units with either the zero/span valve or the IZS option installed, the valve 
control is used. Manifold valves, which may also be present in certain special versions 
of the analyzer, would also use valve control. 
06807C DCN6650