Getting Started 
Teledyne API - T100 UV Fluorescence SO2 Analyzer  
44 
Table 3-5:  Analog Input Pin Assignments 
PIN DESCRIPTION 
DAS 
PARAMETER
1
 
Analog input # 1 
AIN 1 
Analog input # 2 
AIN 2 
Analog input # 3 
AIN 3 
Analog input # 4 
AIN 4 
Analog input # 5 
AIN 5 
Analog input # 6 
AIN 6 
Analog input # 7 
AIN 7 
Analog input # 8 
AIN 8 
GND 
Analog input Ground 
N/A 
1
 See Section 0 for details on setting up the DAS. 
3.3.1.3. CONNECTING ANALOG OUTPUTS  
The T100 is equipped with several analog output channels accessible through a 
connector on the rear panel of the instrument.  The standard configuration for these 
outputs is mVDC.  An optional current loop output is available for each. (Section 
3.3.1.4). 
When the instrument is in its default configuration, channels A1 and A2 output a signal 
that is proportional to the SO
2
 concentration of the sample gas.  Either can be used for 
connecting the analog output signal to a chart recorder or for interfacing with a 
datalogger. 
Output A3 is only used on the T100 if the optional O
2
 or CO
sensor is installed.   
Channel A4 is special.  It can be set by the user (refer to Section 5.9.9) to output any 
one of the parameters accessible through the <TST TST> buttons of the unit’s sample 
display. 
To access these signals attach a strip chart recorder and/or data-logger to the appropriate 
analog output connections on the rear panel of the analyzer.   
 
ANALOG OUT
       A1          A2            A3           A4
   +     -     +     -       +     -      +      -
 
 
Figure 3-7: Analog Output Connector 