5
PWM is the abbreviation of Pulse Width Modulation, and refers to 
the method of controlling signal intensity by controlling the ratio 
between the ON period and OFF period of a pulse signal.
A pulse signal is a signal which repeatedly alternates between ON 
and OFF, and the percentage of ON period during a single cycle is 
referred to as the "duty cycle". 
PWM-controlled lighting is a method for controlling the brightness 
of a lamp by controlling the duty cycle (lit time) of light from a 
pulse-emission source. As the lit time becomes longer, the light 
becomes brighter, and conversely, as the lit time becomes shorter 
the light becomes darker.
Period when LED is lit
Convenient, easy-to-use Excel
®
 add-in software
Reads measurement data from T-10A series Illuminance 
Meters directly into Excel
®
. Further processing of data can 
then be performed easily using the various functions of Excel
®
.
 ⿎
Main specifications of Data Management 
Software T-S10w
Type
Add-in for Excel
®
(Excel
®
 is required to use this add-in.)
Operating 
environment
One of the following environments with Excel
®
 installed:
* Languages in parenthesis (    ) are the OS language.
Windows
®
 XP + Excel
®
 2003 (English, Japanese, or Simplified 
Chinese) 
Windows
®
 7 + Excel
®
 2010 (English, Japanese, or Simplified 
Chinese)
* For details on system requirements for above versions of 
Windows
®
 and/or Excel
®
, refer to their respective specifications.
* Not compatible with 64-bit versions of office 2010.
Compatible  
instruments
T-10A, T-10MA, T-10W
S
A, T-10W
L
A, T-10, T-10M,  
T-10W
S
, T-10W
L
Data Management Software T-S10w (Optional accessory)
About PWM-controlled lighting
Relative Spectral Response
Cosine Correction Characteristics
Mini plug
Mini Cap
T-A14
(for T-10MA)
<
Standard accessories
>
<
Optional accessories
>
AA-size batteries 
2 pcs.
Strap
Case 
T-A10
Cap (with Strap)
T-A13
(for T-10A)
Personal computer 
(Commercial product)
RS-232C printer
(Commercial product)
Data Management software
T-S10w
(Including
T-A15)
Data management using computer
Data recording using printer
USB Cable
T-A15
Printer Cable
T-A12
AC Adapter
External power
T-10A 
Receptor 
Head
Adapter Unit for
Main Body 
T-A20
Adapter Unit for
Receptor Head  
T-A21
With LAN category 5 cable; 1m
For multi-point and cable extension measurement
Example of 
multi-point System
Example of cable
extension
LAN (10 BASE-T) 
category 5 straight cable 
(commercially 
available)
Multi-point measurement requires use of 
optional AC adapter.
Hard Case
CL-A10
Hard Case
CL-A10
Additional receptors (sold separately; Product 
includes 1 receptor according to model purchased)
Custom models
T-10A
T-10MA
T-10W
S
A
T-10W
L
A
Custom order
T-10MA Receptor Head
T-10W
S
A Receptor Head
T-10W
L
Receptor Head
 ⿎
Main Specifications of T-10A
Model
Illuminance Meter T-10A
(Standard receptor head)
Illuminance Meter T-10MA
(Mini receptor head)
Illuminance Meter T-10W
S
A
(Waterproof mini receptor head)
Illuminance Meter T-10W
L
A
(Waterproof mini receptor head)
Type
Multi-function digital illuminance meter with detachable receptor head (Multi-point measurements of 2 to 30 points is possible)
Illuminance meter class
Conforms to requirements for Class AA of JIS C 1609-1: 2006 "Illuminance 
meters Part 1: General measuring instruments" Conforms to DIN 5032 Part 
7 Class B
Conforms to requirements for special Illuminance meters of JIS C 1609-1: 
2006 *
1
Receptor
Silicon photocell
Relative spectral response Within 6% (f
1
´) of the CIE spectral luminous efficiency V (λ)
Cosine response (f
2
)
Within 3%
Within 10%
Measuring range
Auto range (5 manual ranges at the time of analog output)
Measuring function
Illuminance (lx). illuminance difference (lx). illuminance ratio (%). integrated illuminance (lx·h). 
integration time (h). average illuminance (lx).
Measuring 
range
Illuminance 0.01 to 299,900 lx; 0.001 to 29,990 fcd
1.00 to 299,900 lx; 0.1 to 29.990 fcd *
2
Integrated 
illuminance
0.01 to 999,900 x 10
3
 lx·h   0.001 to 99,990 x 10
3
 fcd·h / 0.001 to 9999 h
User calibration function
CCF (Color Correction Factor) setting function: Measurement value x 0.500 to 2.000
Linearity
±2% ±1 digit of displayed value
Temperature/ 
humidity drift
Within ±3%
Computer interface
USB
Printer output
RS-232C
Analog output
1 mV/digit, 3 V at maximum reading; Output impedance: 10 KΩ; 90% response time: 28 ms
Display
3 or 4 Significant-digit LCD with backlight illumination (Automatic illumination)
Power source
2 AA-size batteries /   AC adapter AC-A308  (optional; for 1 to 10 receptors) or 
 
AC adapter AC-A311  (optional; for 1 to 30 receptors)
Battery life
72 hours or longer (when alkaline batteries are used) in continuous measurement
Operating temperature
/humidity range
-10 to 40°C, relative humidity 85% or  less 
(at 35°C) with no condensation 
5 to 40°C, relative humidity of 85% or less 
(at 35°C) with no condensation
Storage temperature /
humidity range
-20 to 55°C, relative humidity 85% or  less 
(at 35°C) with no condensation
0 to 55°C, relative humidity of 85% or less 
(at 35°C) with no condensation
Dimensions
69 x 174 x 35 mm
Main body:  69 x 161.5 x 30 mm
Receptor:  Ø16.5 x 13.8 mm
Cord length
1 m
5 m
10 m
Weight (without battery) 200 g (7.0 oz.)
205 g
260 g (Receptor head only: 120 g)
340 g (Receptor head only: 200 g)
*1 Conforms to requirements for Class AA of JIS C 1609-1: 2006 for all items except cosine response (f
2
).
*2 Although measurements below 1.00 lx are possible, they may not be stable due to the effects of electrical noise.
<Notes regarding mini receptors and waterproof mini receptors> 
*Do not touch the cable during measurements. Doing so may result in unstable measurement values.
*Secure the cable during measurements. Failure to do so may result in unstable measurement values.
Center of 
tripod
socket
Center of 
receptor window 
Center of 
receptor 
window 
7.5
Cord length:
 
T-10MA  ( 1 m)
 
T-10W
S
A  ( 5 m)
 
T-10W
L
A  (10 m)
Ø25
35
174
117.5
161.5
69
Ø45.2
T-10A
T-10MA/ T-10W
S
A/ T-10W
L
A
30
Ø16.5
13.6∼13.9
Ø14
Reference 
plane
Reference 
plane
Ø12.7
Ø3
Ø
6.5
Center of 
tripod
socket
Center of 
receptor window 
Center of 
receptor 
window 
7.5
Cord length:
 
T-10MA  ( 1 m)
 
T-10W
S
A  ( 5 m)
 
T-10W
L
A  (10 m)
Ø25
35
174
117.5
161.5
69
Ø45.2
T-10A
T-10MA/ T-10W
S
A/ T-10W
L
A
30
Ø16.5
13.6∼13.9
Ø14
Reference 
plane
Reference 
plane
Ø12.7
Ø3
Ø
6.5
 ⿎
Dimensions
 (Units: mm)
 ⿎
System diagram
Illuminance Meter T-10A series
Illuminance Meter T-10A series
Ideally, the relative spectral responsivity of the illuminance 
meter should match V (λ) of the human eye for photopic vision.
As shown in the graph above, the relative spectral responsivity 
of Konica Minolta Illuminance Meters T-10A/10MA is within 
6% (f1´) of the CIE spectral luminous efficiency V (λ).
CIE ; Commission Internationale de I´Eclairage
f1´(CIE symbol) ;  The degree to which the relative spectral 
responsivity matches V (λ) is characterized by means of the error f1´. 
Since the brightness at the measurement plane is proportional 
to the cosine of the angle at which the light is incident, the 
response of the receptor must also be proportional to the 
cosine of the incidence angle.
For Konica Minolta Illuminance Meters T-10A/10MA, the 
cosine response f2 is within 3%.
The graph above shows the cosine correction characteristics 
of Konica Minolta Illuminance Meters T-10A/10MA.
Relative sensitivity (%)
Wavelength (nm)
Konica Minolta Illuminance
Meters T-10A/T-10MA
The spectral
luminous efficiency
(λ)
400
450
500
550
600
650
700
750
0
10
20
30
40
50
60
70
80
90
100
60
60
40
40
20
20
0
100
%
80
%
60
%
40
%
20
%
80 
80
Konica Minolta Illuminance
Meters T-10A/10MA
Ideal curve