Average weighted gas flow velocity
Section 6: Meter operation
 Reference, Installation, and Operations Manual
June 2013
3-9000-743 Rev S
In the event of a chord failure with at least one operating chord, the meter’s average weighted 
flow velocity is estimated as shown in 
Equation 6-7  Estimating average flow velocity using proportion values
where
For each non-failed chord, the interpolated proportion value used in 
 is calculated 
as follows:
if the chord’s velocity is surrounded by trained (average velocity, average proportion) 
data pairs, then the interpolated proportion is the linear interpolation between the two 
data pairs
if the chord’s velocity has a trained (average velocity, average proportion) data pair on 
one side of it but not the other, then the interpolated proportion is the data pair average 
proportion
if there are no trained (average velocity, average proportion) data pairs, then the 
interpolated proportion is the corresponding bin’s default average proportion value
Re-acquisition
If all active chords fail, then the meter re-enters the Acquisition mode. If the 
VelHold data point 
is set to a value greater than zero, then, while in the Acquisition mode, the meter holds the 
average weighted flow velocity to the last good value for up to the 
VelHold number of batches 
before setting the velocity to zero. The 
VelHold default value is 0.
=
estimated average weighted flow velocity (m/s) 
(
AvgWtdFlowVel)
=
(non-failed) chord velocity (m/s) 
(
FlowVelA ... FlowVelD)
=
(non-failed) chord interpolated proportion value 
(dimensionless)
V
AvgWtd
est
V
chord
Non
Failed
Chord s
( )
InterpProp
chord
Non
Failed
Chord s
( )
-------------------------------------------------------------
=
V
AvgWtd
est
V
chord
InterpProp
chord