Bacharach AGM300 用户手册

下载
页码 94
Instruction 3015-4275 
iii 
Safety Precautions 
AC Power Supply 
Ensure the source voltage matches the voltage of the product before energizing the equipment. It is highly 
suggested that the AGM300 be placed on a separate circuit (with UPS or surge protection). 
•  A switch or circuit-breaker shall be included in the building installation 
•  It shall be in close proximity to the equipment and within easy reach of the operator 
•  It shall be marked as the disconnecting device for the equipment 
Protective Grounding 
Under no circumstances should this equipment be operated without connection to a protective ground. Doing 
so poses a potential shock hazard and is also a violation of electrical safety standards applicable to this type 
of equipment.  
Explosive Atmosphere 
Do not operate this equipment in the presence of flammable liquids, vapors or aerosols. Operation of any 
electrical instrument in such an environment constitutes a safety hazard. 
Proper Exhaust Venting 
It is imperative that the exhaust port on this instrument be properly vented as described in this manual. Failure 
to do so constitutes a safety hazard. 
Working Inside the Instrument 
Extreme care should be exercised when accessing the interior of this instrument. Only qualified electrical 
maintenance personnel should perform connections and adjustments. Always de-energize the power supply 
before working inside the instrument. 
Modifications to the Instrument 
Under no circumstances should this instrument be modified without written consent from Bacharach, Inc. 
Changes or modifications to this instrument, not expressly approved, will void the warranty. 
In Case of Malfunction 
Do not continue to use this equipment if there are any symptoms of malfunction or failure. In the case of such 
occurrence, de-energize the power supply and contact a qualified repair technician or the nearest Bacharach 
Service Center.  ONLY provided knockouts are to be used for electrical and communication wiring.  Drilling into 
the box will void the warranty.  
Fusing 
AGM300, 120 VAC Input, Power Supply P/N 3015-3167 
F1: 0.5 A, 250V Fast Acting, 5mm x 20mm, P/N 604-2538 
F2: 250 mA, 250V Fast Acting, 5mm x 20mm, P/N 604-2537 
F3: 5 A, 250V Fast Acting, 5mm x 20mm, P/N 604-2539 
F4: 63 mA, 250V Fast Acting, 5mm x 20mm, P/N 604-2535 
 
AGM300, 230 VAC Input, Power Supply P/N 3015-3161 
F1: 250 mA, 250V Fast Acting, 5mm x 20mm, P/N 604-2537 
F2: 125 mA, 250V Fast Acting, 5mm x 20mm, P/N 604-2536 
F3: 5 A, 250V Fast Acting, 5mm x 20mm, P/N 604-2539 
F4: 32 mA, 250V Fast Acting, 5mm x 20mm, P/N 604-2534 
 
ADM800 
F1 & F2: 1 A, 250VAC, Fast Acting, 5mm x 20mm, P/N 3015-4011