Doubler Amp Mode On
Function Group
Menu Map
Description
Programming Codes
See Also
POWER
5
This 
 is applicable to instrument models with a doubler
installed. The doubler has an integral amplifier whose operation is
controlled by the instrument firmware. This 
 turns off the
automatic mode of operation and turns on the amplifier so that it is
always used. This is an unspecified mode of operation since it can
cause increased harmonics and degraded dynamic range at some
frequencies. This 
 has no effect on instruments without a
doubler.
An asterisk next to the key label indicates that this feature is active.
POWer:AMPLifier:STATE 
POWer:AMPLifier:STATE?
Analyzer: 
NONE
Dblr 
 Menu
Dwell Coupled
Function Group
FREQUENCY
Menu Map 
Description
This 
 lets you couple the dwell time for points in the stepped
frequency sweep mode to the ramp sweep mode sweep time. The
equation to determine the dwell time in the dwell coupled mode is as
follows:
Coupled Dwell Time   (sweep time)   (number of step points)
An asterisk next to the key label indicates that this feature is active.
Programming Codes
SCPI: 
SWEep[:FREQ
 
Analyzer: 
NONE
See Also
Step Sup 
Menu
 Operating and Programming Reference
HP 8360
User’s Handbook