HP (Hewlett-Packard) 22A 用户手册

下载
页码 508
 ion : 
Queries the status of any modulation. If any of the modulation states
are on, then it returns a 1, otherwise it returns a 0.
Power Subsystem
Any place where 
 is accepted as a suffix, any level suffix is
accepted also. In the absence of a suffix, the units are assumed to be
as set by the UNIT :POW command.
POWer:ALC:BANDwidthl:BWIDth 
POWer:ALC:BANDwidth?|:BWIDth? 
Sets and queries the ALC bandwidth. This is actually not
continuously variable, so the input is rounded to the nearest possible
switch position. *RST setting is automatically determined since 
is ON.
 
POWer:ALC:BANDwidth|:BWIDth:AUTO?
Sets and queries the automatic ALC bandwidth selection switch. The
 value is ON.
POWer:ALC:CFACtor 
POWer:ALC:CFACtor? 
Sets and queries the coupling factor used when the command
POWer : ALC   
 is set to 
 or 
 
POWer:ALC:SOURce?
Sets and queries the ALC leveling source selection switch.
The *RST value is 
 
POWer : ALC   : 
 ?
Sets and queries the state switch of the ALC. The positions are :
ON-normal ALC operation
OFF-open loop ALC mode
When on, the power can be programmed in fundamental units as
selected by the UNIT:POWer command.
When off, the power is no longer calibrated in absolute units and is
set in units of 
 of arbitrary modulator setting.
POWer:AMPLifier:STATE 
POWer : 
 ier : STATE?
Sets and queries the state of the amplifier contained in
the doubler (for those models with a doubler installed).
 Operating and Programming Reference
HP 8360
User’s Handbook