Model IFT9701 Transmitter Instruction Manual
Getting Started
 continued
Figure 2-1
Location of approvals tag
2.1.2
Orientation and mounting
Orient the transmitter so wiring compartments and conduit openings are easily accessible.
To rotate the transmitter on the sensor manifold or the mounting bracket, use the four supplied 
mounting bolt assemblies. Each bolt assembly includes one M8 bolt, one lock washer, and one 
flat washer. Remove the bolt assemblies, rotate the transmitter, then reinstall the bolt 
assemblies to 12 ft-lb (16 N-m) of torque.
If the transmitter will be integrally mounted to a Micro Motion F-Series sensor, see the sensor 
product data sheet for transmitter and sensor dimensions, and see the instruction manual that 
was shipped with the sensor for information about flowmeter mounting and location.
If the transmitter will be remotely mounted, use the supplied bolt assemblies to attach the 
transmitter to the mounting bracket. Attach the bracket to a rigid, stable surface or instrument 
pole that will not transfer excessive vibration into the transmitter. See Chapter 3 for more 
information about remote mounting.
2.1.3
Temperature, humidity, and vibration
Install the transmitter according to specified limits:
Ambient temperature
-
without optional LCD: –22 to +131 °F (–30 to +55 °C)
-
with optional LCD: 32 to 131 °F (0 to 55 °C)
Humidity: 5 to 95% non-condensing
Vibration: Meets IEC 68.2.6, 2 g
Hazardous area 
approvals tag