OMWT-SOFT-OPC Server
OMWT-SOFT-OPC Main Window x 9
Add/Delete IP Connection
(OMWT-RECIP Receiver only)
OWMT-SOFT-OPC will display the “Add/Delete IP Connection” screen.  IP 
addresses already setup will be displayed.  The following is a list of the columns:
Node – Node number for the IP address.  OWMT-SOFT-OPC uses this 
number to identify this IP connection.  OWMT-SOFT-OPC uses this 
number to update the OPC item “nodaladdress”.
IP Address – the IP address that is used to address the receiver.
Port – the port number that is used along with the IP address to address the 
receiver.
Conn – identifies who initiated the connection.  Outgoing: OWMT-SOFT-
OPC initiated the connection with the receiver.  Incoming: the receiver 
initiated the connection with OWMT-SOFT-OPC.
Password – initial password used to gain access to the receiver.
Description – a general description field associated with the IP connection.
The following is list of buttons in the “Add/Delete IP Connection” screen:
Test – OWMT-SOFT-OPC will try to make a connection to this IP address 
and report back success or failure.
Add New – OWMT-SOFT-OPC displays the “New IP Address” screen.  
The following is list of the fields to enter to add a new IP connection:
IP Address – the IP address that is used to address the receiver
Port – the port number that is used along with the IP address to address 
the receiver.
Description– a general description field associated with the IP 
connection.
Password – initial password used to gain access to the receiver.  If the 
receiver does not have a password set or does not require a 
password then leave this field blank.
Connection – identifies whether OWMT-SOFT-OPC will initiate the 
connection or whether the receiver will initiate the connection.  
Outgoing: OWMT-SOFT-OPC initiates the connection.  Incoming: 
OWMT-SOFT-OPC waits for the receiver to initiate the 
connection.
Edit – OWMT-SOFT-OPC displays the “Edit” screen showing the selected 
IP connection.  The “Edit” screen has the same fields as the “New IP
Address” screen.
Delete – delete the selected IP connection.