OMWT-SOFT-OPC Server
OMWT-SOFT-OPC Server .ini File Format x 23
ALARMTEMP
(OMWT-TEMPALRM)
[Devicen] - Section name.
Address - serial number of the sensor.
DeviceType - Device type. (DeviceType=TEMP)
PollTime - How often to poll the device. (ms)
Label - Name used for the device (and Topic).
Log - Enable/disable logging of device.
LogDecimalPlaces - Number of decimals to create when logging.
Units - See OPC item temp.input.units for description.
Cal1Raw - Two point calibration point 1.
Cal2Raw - Two point calibration point 2.
Cal1Engr - Two point calibration engineering units for point 1.
Cal2Engr - Two point calibration engineering units for point 2.
DeadBand - Dead band filtering value in ºC. (Default 0.01)
FilterElements - running macro filter of the sampled data.  Syntax: [no. of 
elements, type] where no. of elements is number of samples to filter; 
type – “AVERAGE”, “MEDIAN” and “AVERAGEMEDIAN”.  
Default is “=3,MEDIAN”.  Example: FilterElements1=6,AVERAGE
Description - General purpose description.
CorrID – What CorrectionID to use with this device.  Must be found in the 
"Correction.INI" file that you write and keep in the same directory as 
the OMWT-SOFT-OPC.ini file.  0 is the default and means no 
correction; 127 is the maximum