A&D MX-50 用户手册

下载
页码 75
MS-70 / MX-50 / MF-50 / ML-50 
34 
 
 
Store Parameters and Finish the Operation
Store Parameters and Finish the Operation
Store Parameters and Finish the Operation
Store Parameters and Finish the Operation    
 14.  Press the 
 ENTER 
  key to store the new parameters of 
the measurement program to program number 3. 
Pressing the key, the weighing mode is automatically 
displayed. 
When PROG 3 is recalled, the settings can be used. 
 
 
  To cancel the new parameters and return to the 
weighing mode, press the 
 RESET 
 key. 
 
 
 
Timer Mode
Timer Mode
Timer Mode
Timer Mode    
Refer to page 32 for Standard Mode, Automatic Mode or Manual Mode 
 
 
  This explanation uses the following parameters and sample displays of MX-50. 
 
 
 
 
 
 
 
  
Common 
Items 
 
   
Program number.......................................... 4    ( PROG 4 ) 
 
   
Analysis mode ............................................. Timer mode (Symbol:    U-t 
)
 
   
Drying program 
Heating pattern ....................Ramp drying (
 
  ) 
 
   
 
Final drying temperature .....160°C 
 
   
 
Ramp time............................5.0 minutes 
 
   
 
Drying time...........................10.0 minutes 
  
 Measurement 
unit ...............................................Moisture 
content 
 
   
Minimum scale value during measurement.. 0.01 % 
 
   
Minimum scale value of gram display .......... 0.001 g 
 
   
Sample quantity ........................................... Approximately 5 g 
 
   
Data memory function.................................. Not used 
 
 
 
    
 
1.  Display gram unit (of the weighing mode). 
 
Select a Program Number to Edit the Measurement Program
Select a Program Number to Edit the Measurement Program
Select a Program Number to Edit the Measurement Program
Select a Program Number to Edit the Measurement Program    
 2. Press 
the 
 PROGRAM 
  key and press the 
 or 
 key to select a program number. 
 
 3. Press 
the 
 ENTER 
  key to use the number. 
 
 
4.  The analyzer displays 
 end 
  and returns to the 
weighing mode. 
 
Caution 
If the data memory function is active, the data 
memory number (MEM) is displayed in place of the 
program number (PROG).
 
% MOIST
/W
/D
PROG 1
 0.000
g
 
PROG "4
 
 
 end
 
PROG 4
 0.000
g
 
Program number 
Final drying temperature
50°C 
Ramp time 
Drying time
 0.000
g
 
160°C 
Gram display of 
weighing mode