Honeywell 96D 用户手册

下载
页码 80
M-511401
  96d User Manual 
61
8/10
Honeywell 
System Programming
Network Menu
Remote Calibration
The network menu also offers a Calibration option for use with devices 
that support network calibration. 
The Calibration screen contains four (4) lines of information:
Line 1: 
Indicates the mode (Calib, meaning calibration), the 
(Modbus) address of the device to calibrate (001) and the 
type of device to calibrate (301D2)
Line 2: 
Indicates the status (Normal or In calib…) of the specified 
device
Line 3: 
Displays the function to perform (Set Zero)
Line 4: 
Displays the function to perform (Set Span) and the span 
gas concentration value (246 ppm)
5.
On the first line, scroll to the device address and press Enter
6.
Scroll through the devices to display the desired device* and press 
Enter to select.
7.
The second line displays the device’s status
8.
Scroll to select the desired function, Set Zero to set the device's 
zero, and press Enter to select.
9.
Upon pressing Set Zero, the controller requests confirmation.  
*The device must be configured in the 96D’s database in order to be 
included in the device addresses displayed on screen.
7. Network
Statistics
Calibration
Calib 007
S301M
Status:
Normal
 
Set Zero
 
Set Span
246
PPM
Calib 007
S301M
Status:
Normal
 
Set Zero
 
Set Span
246
PPM