SCXI-1125 Calibration Procedure
Specifications
The following table contains the specifications you should use when
verifying and calibrating your SCXI-1125 module. If your device has been
calibrated within the last year, the test point value should fall between the
Upper Limit and Lower Limit values.
Table 1. SCXI-1125 Specifications at 25 °C ±10 °C
Gain
Range
Test Point
Upper Limit (V)
Lower Limit (V)
1
5
4.950000
4.9623181
4.9376819
1
0
0.000000
0.005002
–0.005002
1
–5
–4.950000
–4.9376819
–4.9623181
2
2.5
2.475000
2.4811601
2.4688400
2
0
0.000000
0.002502
–0.002502
2
–2.5
–2.475000
–2.4688400
–2.4811601
5
1
0.990000
0.9924652
0.9875348
5
0
0.000000
0.001002
–0.001002
5
–1
–0.990000
–0.9875348
–0.9924652
10
0.5
0.495000
0.4962336
0.4937664
10
0
0.000000
0.000502
–0.000502
10
–0.5
–0.495000
–0.4937664
–0.4962336
20
0.25
0.247500
0.24811781
0.24688220
20
0
0.000000
0.000252
–0.000252
20
–0.25
–0.247500
–0.24688220
–0.24811781
50
0.1
0.099000
0.09924832
0.09875168
50
0
0.000000
0.000102
–0.000102
50
–0.1
–0.099000
–0.09875168
–0.09924832
100
0.05
0.0495000
0.04962516
0.04937484
100
0
0.000000
0.000052
–0.000052
100
–0.05
–0.049500
–0.04937484
–0.04962516
200
0.025
0.024750
0.02481358
0.02468642
200
0
0.000000
0.000027
–0.000027